By Kurnia · Published November 11, 2026 · Version 1 · 20 min read
Medical devices from the building technician’s side
power supply and outlets, protection class and leakage current, electrical safety testing, calibration by authorized bodies, inventory and fault reporting, and limits of authority
A nurse feels a tingle when touching a patient’s bed frame, and the building technician finds a cracked multi-outlet adapter under the bed supplying four devices at once. This chapter covers medical devices from the side of the person who looks after the building: the power supply and outlets that devices are plugged into, protection class and where leakage current comes from, why the leakage currents of several devices add up on one protective conductor, electrical safety testing of devices and who carries it out, testing and calibration by authorized institutions, inventory, maintenance, and fault reporting, and the limits of the building technician’s authority relative to electromedical technicians, medical device maintenance units, testing institutions, and medical staff. The status of the testing and calibration rules as of November 2026 is stated with its source, including the revocation of Permenkes 54/2015.
This material involves real hazards. A concept explanation doesn’t replace site procedures, competency, energy isolation, the latest SDS, or a work permit.
Shortly before the morning shift change on the internal medicine ward of a hospital, a nurse felt a tingle in their hand while holding the metal frame of a patient’s bed. The building technician who was called did not touch anything at first. They saw that under the bed there was a multi-outlet adapter with a cracked housing supplying four devices at once: the bed motor, an infusion pump, a mattress warmer, and a phone charger belonging to the patient’s family. The question is not only which outlet is faulty, but also who may touch what: the wall outlet belongs to the building installation, the infusion pump belongs to the medical device maintenance unit, and the decision to move the patient or unplug equipment belongs to the medical staff.
Scope
This chapter covers electrically powered medical devices from the building technician’s side: the power supply and outlets that devices are plugged into, protection class and leakage current, electrical safety testing of devices, testing and calibration by authorized institutions, inventory and fault reporting, and limits of authority. The fixed installation of medical rooms (room groups, medical IT systems, alternative supply, equipotential bonding) is covered in Chapter 102 on electricity in medical rooms. The effects of current on the body are in Chapter 75 on electrical hazards to the body, equipment classes and shock protection in Chapter 21, insulation resistance testing in Chapter 17, grounding and RCD testing in Chapter 18, final circuits and socket-outlets in Chapter 32, and calibration and traceability of measuring instruments in Chapter 61. The hospital facility maintenance program as a whole is in Chapter 31 · Buildings & Facilities on hospital facility and utility management.
This chapter gives no steps for opening, repairing, adjusting, or testing medical devices, gives no method for measuring device leakage current, and does not cover the use of devices on patients. Repair and testing of devices belong to electromedical technicians and the medical device maintenance unit, periodic testing and calibration belong to the authorized testing institution, and clinical decisions belong to the medical staff. The building technician looks after the installation up to the outlet, recognizes danger signs, records them, and reports.
Medical devices at the end of the outlet: who owns what
Medical device (Indonesian alat kesehatan) is a material, instrument, apparatus, machine, and/or implant that does not contain a drug and is used to prevent, diagnose, cure, and relieve disease, care for the sick, restore health, and/or form structures and improve body functions. This chapter limits itself to medical devices that use electricity and are plugged into the building installation, for example infusion pumps, patient monitors, ventilators, motorized beds, and patient warmers.
In an ordinary building the boundary of responsibility is clear: the installation up to the outlet belongs to the electrician, and the equipment behind the plug belongs to its owner. In a hospital that line still exists, but the two sides depend on each other. A class I medical device is only safe if the protective conductor of the outlet is intact, and a sound outlet does not help if the device cable is cracked. The regulation on medical device maintenance allows the medical device maintenance function to be combined with the maintenance of facilities and infrastructure or to stand on its own. In a combined unit, the person responsible for infrastructure coordinates electricity, water, gas, steam, and air conditioning, while the person responsible for medical equipment coordinates the devices.
For the building technician, this division decides the first step in every event. Faults on the wall side (outlet, circuit, panel) are reported to the person responsible for infrastructure. Faults on the device side (power cord, plug, device housing, device alarm) are reported to the person responsible for medical equipment. The patient’s condition is reported to the medical staff on duty.
Power supply and outlets
Electronic medical devices are sensitive to power quality. Ideally the voltage is steady at the right value, free of spikes, surges, and dropouts, and rarely lost; the WHO maintenance guidance advises device technicians to work with those who manage the electrical system, for example with voltage regulators, UPS units, and surge protectors, to avoid extension leads or plug boards connected in series, and to make sure the backup generator works and its changeover takes less than 10 seconds. The stricter changeover time for group 2 rooms is covered in Chapter 102.
Outlets in patient rooms carry several markings that the building technician needs to recognize. According to Permenkes (Minister of Health Regulation) 2306/2011, which as of November 2026 has been revoked (Chapter 102), where the supply voltage, current, or frequency differs, the plugs used must not be interchangeable, and socket-outlets supplied from the safety-services source must be easy to identify. Outlets on the medical IT system are made so that they cannot be used on another system or are marked clearly and permanently. These markings tell staff which devices stay on when the normal supply is lost, so a marking that has come off or been swapped is reported like a fault.
Multi-outlet adapters and extension cord reels are the problem building technicians most often find on wards. The UK guidance on managing medical devices states that outlet adapters should not be used on medical devices. The reasons are electrical and mechanical. All devices on one multi-outlet adapter share one protective conductor path, one plug contact, and one cable; if any of these breaks or works loose, every class I device behind it loses protection at once. Load current also adds up, and leakage current adds up too, as discussed in the next section.
The load on a circuit is calculated from the sum of the currents of the devices running at the same time:
where is the circuit current in amperes, is the power of device in watts, and is the supply voltage in volts. This formula assumes a power factor close to one and does not include the momentarily larger starting current of motors or heaters. Device power and voltage data are read from the device nameplate or from the inventory record, which does include the power requirements of each device.
Protection class and leakage current
Equipment class is the classification of a device according to how it protects against electric shock if its basic insulation fails, as discussed in Chapter 21 on shock protection. A class I device relies on basic insulation plus a connection of the device enclosure to the protective conductor, so its safety depends on the protective contact of the plug and outlet. A class II device relies on double or reinforced insulation and does not use a protective conductor. The nameplate or label of the device shows its electrical class, and recognizing electrical class symbols is among the skills expected of staff who check devices on delivery.
Leakage current is current that flows from the live parts of a device to the device enclosure, to ground, or to the patient through insulation and capacitance, even when the device is not faulty. One of its origins is the capacitance between live conductors and the metal enclosure, for example interference-suppression capacitors connected from phase to the protective conductor. For a purely capacitive component,
where is the leakage current in amperes, is the frequency in hertz, is the capacitance in farads, and is the voltage in volts. This formula applies only to the capacitive part; damp or damaged insulation adds a resistive component.
In a healthy class I device, the leakage current to the enclosure is carried by the protective conductor to ground and is not felt. If the protective conductor is broken in the device cable, in the plug, or in a multi-outlet adapter, the same current looks for another path, for example through the hand of a nurse touching the device enclosure and a grounded bed frame. That is the most likely explanation of the opening case. For a patient connected to a catheter or electrode near the heart, a far smaller leakage current is already dangerous, as explained in Chapter 102.
That is why the WHO maintenance guidance calls for ground resistance and leakage currents to be measured after maintenance, especially after procedures that may have affected the device’s safety features, to make sure the values are within the applicable limits. Those limits are set by the medical device safety standards and the manufacturer. This chapter does not give the figures; the building technician reads them from the test result sheet issued by the tester, not from memory.
Electrical safety testing of devices
Electrical safety testing of a medical device is different from testing the installation. Installation testing checks the building’s wiring, panels, grounding, and RCDs (Chapter 17 and Chapter 18). Device testing checks the device itself: its physical condition, ground resistance, and leakage current, followed by the device’s function. Every sample maintenance procedure in the WHO guidance includes a step of inspecting the power cord, strain relief, and plug for signs of damage.
The UK guidance distinguishes two kinds of test. Type tests under product safety standards are designed for type testing and can damage the device, so some of them are not suitable for pre-use or maintenance tests. Acceptance tests must not exceed the bounds of normal use, and for periodic tests and tests after repair that guidance refers to BS EN 62353:2014 as best practice. When a device is delivered, the same guidance advises measuring and recording “as new” reference values for electrical safety parameters as a basis for comparison at later maintenance.
The test equipment is specialized as well. The list of test equipment in the WHO guidance names an electrical safety analyzer for all electrical equipment, receptacle testers for electrical outlets, and an isolated power tester for IT systems. This split reflects the division of work: receptacle testers and isolated power testers sit on the building installation side, while the device electrical safety analyzer is in the hands of the electromedical technician or the testing institution.
Electrical safety testing of a device is carried out on delivery, after repair, and periodically. The building technician does not carry it out. The building technician’s role is to make sure the outlets where the device is tested and used are in good condition, hand over suspect devices, and record installation findings that may explain a test result.
Testing and calibration by authorized bodies
Calibration is the activity of verification to determine the correctness of the indicated value of a measuring instrument and/or measuring material. Testing is the act of physical inspection and measurement to compare a device with a standard or to determine a quantity or a measurement error. For hospitals, the government regulation implementing the health law states that health equipment must meet service standards, quality, security, and safety requirements and be fit for use, and must be tested and calibrated periodically by the authorized health facility testing institution.
The details used to be set by Permenkes 54/2015 tentang Pengujian dan Kalibrasi Alat Kesehatan (on testing and calibration of medical devices). That regulation set device testing as covering function tests, safety tests, and performance tests; testing and calibration at least once a year; and earlier testing, among other cases, when the output or safety no longer conformed, after repair, after relocation of devices that require installation, or when the device had no certificate yet. Devices that met the standard were given a certificate and a fit-for-use label, devices that did not were given a statement and an unfit-for-use label, and the label shows, among other things, the device serial number, the test date, the statement of fit or unfit, and the deadline for retesting. As of November 2026, JDIH (legal documentation and information network) Kemenkes records Permenkes 54/2015 as no longer in force because it was revoked by Permenkes 5/2026 tentang Perbekalan Kesehatan (on health supplies); no replacement provision detailing intervals and labels has been found, so the applicable interval is checked with the hospital’s device maintenance unit and the testing institution.
Among those institutions, the technical implementation units of the Ministry of Health for the safety of medical devices and health facilities carry out testing and/or calibration of medical devices, calibration of standard measuring instruments, and inspection of health facility infrastructure. Calibration requires traceability to standards, as discussed in Chapter 61 on instrument calibration and traceability, and therefore cannot be replaced by “checking against another device”.
The building technician does not test or calibrate medical devices. What they can do is read the labels: devices whose label states unfit for use, whose retest deadline has passed, or whose label is missing are recorded and reported to the person responsible for medical equipment.
Inventory, maintenance, and fault reporting
Health service facilities must maintain their medical devices, and maintenance activities cover inventory, promotive maintenance, function monitoring or inspection, preventive maintenance, and corrective maintenance or repair. Medical device inventory is the collection, recording, and reporting of device data based on investment value, technical life, risk, or a combination of these; devices with a technical life of less than one year should not be inventoried. The maintenance provider must record and report the type, number, and condition of the devices maintained every six months through a Ministry of Health application, and the facility provides a maintenance budget of at least 4 percent of the asset value of its medical devices per year. As of November 2026, JDIH Kemenkes records this Permenkes 15/2023 as in force.
A good inventory record contains a unique identification number, the type of device, the manufacturer, the model, the serial number, the location, the condition or operating status (in service or not, with the reason), the power requirements, and the maintenance service provider. The power requirements column matters to the building technician, because it shows which devices need a different voltage, three phases, or special attention before being moved to another room.
Not all devices are treated alike. The WHO guidance includes a sample risk-based program that gives each device type an equipment management number, the sum of scores for function, physical risk of use, maintenance requirements, and failure history:
where is the device function score (life support 10, down to miscellaneous 2), is the score for the physical risk of use (potential patient death 5, down to no significant risk 1), is the maintenance requirement score (extensive 5, down to minimal 1), and is the failure history factor from plus 2 to minus 2. In that sample, devices with an EM of 12 or more are included in the inspection and preventive maintenance program, an EM of 15 or more is inspected at least every six months, an EM of 19 or 20 every four months, and devices with extensive maintenance requirements (score 4 or 5) receive preventive maintenance every six months. These scores are a sample from the guidance, not a national rule.
For the building technician, fault reporting starts with the label. When a problem is found and the device must not be used, the WHO guidance asks the maintenance technician to raise a repair request, attach a label stating that the device is out of service, and inform the user department. The UK guidance asks for every event that causes or could cause unwanted effects on the safety of patients, users, or other people to be reported, including near misses. The tingle in the opening case is such an event even though no one was hurt; incident investigation is covered in Chapter 12 · Safety.
Limits of authority
The Health Law 17/2023 classifies electromedical technicians as health professionals in the biomedical engineering group. Medical device maintenance providers, whether a hospital-owned unit, a regional-government unit, or a business entity, must have staff with medical device maintenance competence proven by a training certificate from an accredited training institution. A hospital-owned maintenance unit only maintains the devices of that hospital.
The division of work that follows from these rules:
- Building technicians look after the installation up to the outlet, recognize multi-outlet adapters, damaged cables, and expired labels, record, and report.
- Electromedical technicians and the medical device maintenance unit inspect, maintain, repair, and electrically safety-test devices, supported by the device manufacturer or supplier when needed.
- The authorized health facility testing institution tests and calibrates devices periodically and issues certificates and labels.
- Medical staff decide whether a device may be disconnected from a patient, replaced, or moved, because that decision concerns the patient’s condition.
This division also means that the building technician does not unplug a device that is connected to a patient, does not open a device housing, does not replace fuses inside a device, does not bypass device alarms, and does not move test labels.
Worked example
The numbers in the examples below are practice numbers, unless their origin is stated.
Example 1: leakage current of a suppression capacitor. A device has a total capacitance of 10 nF from its live conductors to the enclosure. At 230 V and 50 Hz, its leakage current is A, that is, 0.72 mA. As long as the protective conductor is intact, that current flows to ground without being felt. If the protective conductor is broken, the same current can pass through the body of a person touching the device enclosure and a grounded object.
Example 2: leakage currents adding up on a multi-outlet adapter. Four class I devices on one multi-outlet adapter each leak 0.3 mA. The protective conductor of the adapter carries mA. If the protective contact of the adapter breaks, the four device enclosures are connected to each other but not to ground, and the 1.2 mA current looks for a path through a person touching one of them. A current of that size is above the 0.5 mA perception line in Chapter 75, so the opening case makes sense even without a truly faulty device.
Example 3: load on a bed circuit. One 16 A circuit in a ward supplies a 1,200 W patient warmer, 3 infusion pumps of 25 W each, a 60 W monitor, and a 150 W air mattress at 230 V. The current is A. If the patient’s family adds a 2,000 W water heater, the current becomes A, very close to 16 A, not counting starting current. An MCB that trips because of this load cuts off the infusion pumps at the same bed.
Example 4: the minimum maintenance budget. The asset value of a hospital’s medical devices is 15 billion rupiah. The minimum maintenance budget is billion rupiah per year, that is, 600 million rupiah. The maintenance report is submitted twice a year, because .
Example 5: equipment management number. Ventilator: life-support function 10, risk of patient death 5, extensive maintenance 5, average history 0, so . This device is inspected every four months, that is, times a year, and receives preventive maintenance every six months. An infusion pump with scores 9, 4, 3, and a poor history of plus 1 gets , is inspected at least every six months, and because its maintenance score is 3 it receives preventive maintenance annually. The function and risk scores of these two devices are practice assessments, not official classifications.
Example 6: reading labels on a round. On one floor, the building technician records 120 devices with test labels. Seven labels are past their retest deadline, three labels state unfit for use, and two devices have no label. Reported to the person responsible for medical equipment are devices, that is, percent of the devices on that floor. The building technician does not withdraw devices in use on patients; they hand over the list and inform the head of the unit.
Stop conditions
- Someone feels a shock or a tingle when touching a medical device, a bed, or a metal object near a patient: do not touch it again and do not unplug a device connected to a patient; ask the medical staff to deal with the patient and the device, report to the electromedical technician and the person responsible for infrastructure, and mark the outlet in use.
- An outlet, plug, or device power cord looks scorched, cracked, hot, or loose: stop using an outlet that is not supplying patient equipment; if it is supplying patient equipment, the medical staff move the equipment first, then the outlet is repaired by an authorized electrical technician and the device is checked by the electromedical technician.
- A device is labeled unfit for use, its retest deadline has passed, or its label is missing: do not use it for a new patient and do not move its label; report to the person responsible for medical equipment.
- The building technician is asked to open a device housing, replace a fuse inside a device, repair a device cable, or bypass a device alarm: refuse and hand it over to the electromedical technician or the medical device maintenance unit.
- The building technician is asked to measure leakage current or carry out an electrical safety test on a device, especially a device connected to a patient: refuse; device testing is done by an electromedical technician or the authorized testing institution.
- A device has been dropped, has had liquid spilled on it, or has come back from repair without a test record: mark it do not use and hand it to the medical device maintenance unit before it is used again.
- A multi-outlet adapter, extension cord reel, or plug adapter is found supplying medical devices: do not unplug it yourself if the device is in use on a patient; report to the head of the unit and the electromedical technician so that the device is moved to a wall outlet.
- The MCB of a medical device circuit trips repeatedly: do not raise the rating of the protective device and do not switch it on again and again; report to the person responsible for the installation and the electromedical technician.
Common misinterpretations
Guess what’s wrong first, then open the explanation.
Common misconception
A device from the factory is already safe, so it does not need testing.
Why is it wrong?
Devices are tested on delivery, after repair, and periodically, because cables, plugs, and insulation change during use, and hospital health equipment must be tested and calibrated periodically by the authorized institution.
Common misconception
A good multi-outlet adapter may be used for medical devices.
Why is it wrong?
The UK guidance rejects outlet adapters for medical devices. All devices on one multi-outlet adapter share one protective conductor, so one broken contact removes the protection of every device and their leakage currents add up.
Common misconception
Leakage current means the device is faulty.
Why is it wrong?
Every healthy device leaks a small current through its capacitance. What is dangerous is leakage current that exceeds the limit or that has lost its path to the protective conductor.
Common misconception
Calibration can be done by comparing two identical devices.
Why is it wrong?
Calibration requires traceability to standards and is done by an authorized institution that issues a certificate and label.
Common misconception
The building technician may unplug a device that is dangerous.
Why is it wrong?
Unplugging a device connected to a patient is a clinical decision. The building technician stops the use of an outlet that is not supplying a patient and asks the medical staff to move the connected device.
Links to other chapters
Chapter 102 on electricity in medical rooms covers the fixed installation that these devices are plugged into, including medical IT systems and equipotential bonding. Chapter 21 on shock protection explains equipment classes, Chapter 75 the effects of current on the body, and Chapter 61 calibration traceability. Chapter 31 · Buildings & Facilities on hospital facility and utility management places device inventory and maintenance within the facility program, and Chapter 12 · Safety covers incident and near-miss investigation.
Exercises and worked solutions
The numbers in the exercises are for practice.
Try to answer first, then open the worked solution.
Calculate the capacitive leakage current of a device with a capacitance of 4.7 nF at 230 V and 50 Hz.
Check my answer
See the worked solution
A, that is, 0.34 mA. This current is normal for a healthy device and safe as long as its protective conductor is intact, because the current flows to ground through that conductor.
Three devices on one extension cord reel each leak 0.25 mA. What is the current in the protective conductor of the reel, and what happens if that conductor breaks?
Check my answer
See the worked solution
The current is mA, because all three share one protective conductor. If that conductor breaks, all three device enclosures lose their grounding at once, and the current looks for a path through a person touching a device enclosure and a grounded object.
A patient’s family asks the building technician to “fix” an infusion pump whose alarm keeps sounding. What is the technician’s answer?
Check my answer
See the worked solution
An infusion pump is a medical device repaired by an electromedical technician or the medical device maintenance unit, so the building technician does not open it or silence its alarm. The building technician informs the nurse on duty and reports to the person responsible for medical equipment.
The asset value of a hospital’s medical devices is 8 billion rupiah. What is the minimum maintenance budget per year?
Check my answer
See the worked solution
billion rupiah, that is, 320 million rupiah per year. This figure is a lower limit; the actual budget depends on the number of high-risk devices and their maintenance contracts.
A patient monitor is repaired by the vendor and returned to the ward without a test record. What must be done?
Check my answer
See the worked solution
A repair can change the device’s safety features, so the device is tested after repair before it is used again. The monitor is marked do not use and handed to the medical device maintenance unit until there is a test record.
Why is the power requirements column in the inventory record useful to the building technician?
Check my answer
See the worked solution
That column shows the voltage, phases, and special requirements of the device, so the building technician can make sure the outlets and circuits in the destination room are suitable before the device is moved. Without that data, a device can be plugged into an unsuitable outlet or overload a circuit that is already full.
References
External sources
- Peraturan Menteri Kesehatan Nomor 15 Tahun 2023 tentang Pemeliharaan Alat Kesehatan di Fasilitas Pelayanan Kesehatanjdih.kemkes.go.id
- Peraturan Pemerintah Nomor 28 Tahun 2024 tentang Peraturan Pelaksanaan Undang-Undang Nomor 17 Tahun 2023 tentang Kesehatanjdih.kemkes.go.id
- Peraturan Menteri Kesehatan Nomor 54 Tahun 2015 tentang Pengujian dan Kalibrasi Alat Kesehatanjdih.kemkes.go.id
- Peraturan Menteri Kesehatan Nomor 28 Tahun 2023 tentang Organisasi dan Tata Kerja Unit Pelaksana Teknis Bidang Pengamanan Alat dan Fasilitas Kesehatanjdih.kemkes.go.id
- Undang-Undang Nomor 17 Tahun 2023 tentang Kesehatanjdih.kemkes.go.id
- Medical equipment maintenance programme overviewiris.who.int
- Introduction to medical equipment inventory managementiris.who.int
- Managing Medical Devices: Guidance for health and social care organisationsassets.publishing.service.gov.uk
- Peraturan Menteri Kesehatan Nomor 2306/MENKES/PER/XI/2011 tentang Persyaratan Teknis Prasarana Instalasi Elektrikal Rumah Sakitjdih.kemkes.go.id
Revision history
| Date | Version | Change |
|---|---|---|
| November 11, 2026 | 1 | First published |
How to cite
Kurnia (2026). Medical devices from the building technician’s side: power supply and outlets, protection class and leakage current, electrical safety testing, calibration by authorized bodies, inventory and fault reporting, and limits of authority. ARSTEMA Academy. Revised November 11, 2026. arstema.com/en/academy/konsep/alat-medis-dari-sisi-teknisi-gedung-catu-daya-dan-stopkontak-kelas-proteksi-dan-arus-bocor-uji-keselamatan-listrik-kalibrasi-oleh-pihak-berwenang-inventaris-dan-pelaporan-kerusakan-serta-batas-kewenangan
Self-check · 3 questions
Check your understanding before moving on
Write your answer in your own words, then compare it with the worked solution from this chapter’s exercises. The system doesn’t judge open answers as right or wrong; you decide which parts you’ve understood.
Safety limits still apply. This quiz tests reasoning; it doesn’t authorize mixing, adjusting, dismantling, live measurement, or field action.
Calculate the capacitive leakage current of a device with a capacitance of 4.7 nF at 230 V and 50 Hz.
Three devices on one extension cord reel each leak 0.25 mA. What is the current in the protective conductor of the reel, and what happens if that conductor breaks?
A patient’s family asks the building technician to “fix” an infusion pump whose alarm keeps sounding. What is the technician’s answer?
Nearby chapters
Reading order by shelf: Electrical & Control · Chapter 104.
Reading order by shelf: Electrical & Control · Chapter 102.