AEDs and first aid boxes from the facility side
placement and signs, checking batteries and electrodes, expiry dates, first aid box contents, record keeping, and the role of trained officers
A hotel guest collapses in the fitness center, and the AED brought from the lobby turns out to be showing a warning because its electrodes expired five months earlier, while the nearest first aid box holds headache tablets but has run out of sterile gauze. This chapter covers AEDs and first aid boxes from the facility side: the duties for facilities and P3K (first aid) officers under Permenakertrans PER.15/MEN/VIII/2008, how an AED guides its user, AED placement, cabinets, and signs, routine checks of the status indicator, batteries and electrodes and their expiry dates, the contents and number of first aid boxes under that regulation’s annexes, record keeping, and the division of roles between building technicians, licensed P3K officers, suppliers, and emergency services. The AED maintenance provisions come from guidance by the UK Department for Education and by HSE, and are presented as UK guidance. This chapter does not teach medical procedures; the only actions discussed near a casualty are calling trained officers and emergency services, bringing the AED, and following the device’s voice prompts.
One afternoon, a guest collapsed on a treadmill in the fitness center on floor 5 of a hotel. The instructor called the P3K (first aid) officer on duty and phoned emergency medical services, while a security officer ran to the lobby to take the AED from its cabinet. The round trip took almost four minutes because the lift was in use by a tour group. When the AED was opened, it switched on and gave voice prompts, but its status indicator had been red for several weeks, and the electrode label showed an expiry date five months earlier. The first aid box in the fitness room did not help either: the sterile gauze had run out, while inside were headache tablets and an ointment that nobody had recorded. After the incident, the engineering team was asked to take over the care of the AED and the first aid boxes, and the technician needs to know the right locations, what is checked and how often, when batteries and electrodes are replaced, what the regulation says a first aid box contains, and the limits of the job.
Scope
This chapter covers AEDs and first aid boxes from the facility side: the regulatory basis for first aid facilities and P3K officers, how an AED guides its user as far as a facility manager needs to know, placement, cabinets, and signs, routine checks, batteries, electrodes, and expiry dates, the contents and number of first aid boxes, record keeping, and the division of roles.
This chapter does not teach medical procedures. First aid, cardiopulmonary resuscitation, and how to apply electrodes to a casualty are learned through official P3K officer training. The only actions near a casualty mentioned here are calling a trained P3K officer and emergency medical services, bringing the AED to the scene, and following the device’s voice prompts. Electrical hazards to the body and their first aid are covered in Chapter 75 · Electrical & Control, chemicals and their first aid in Chapter 6, personal protective equipment in Chapter 5, hotel fitness equipment in Chapter 17 · Mechanical & Mechatronics, and fire evacuation drills in the R23 chapter on fire evacuation drills in hotels (Chapter 14).
Outside the building technician’s authority: giving medical treatment other than the actions above, setting the workplace hazard classification, opening the AED casing, repairing the AED or its battery, updating its software, downloading the recorded event data, and installing the power supply for an outdoor AED cabinet. That work belongs to licensed P3K officers, company management, the AED supplier or manufacturer, emergency medical services, and authorized electrical technicians.
Regulatory basis: first aid facilities and officers in the workplace
Workplace P3K (first aid at work), under Permenakertrans (Minister of Manpower and Transmigration Regulation) PER.15/MEN/VIII/2008, is the effort to give first aid quickly and correctly to workers and other people present at the workplace who become ill or injured at the workplace. For a hotel, “other people” includes guests, event visitors, and contractors. The employer must provide P3K officers and P3K facilities, and the management must carry out P3K at the workplace. As of November 2026, JDIH Kemnaker lists this regulation as in force.
A P3K officer is a worker appointed by the management and given the additional task of carrying out P3K. The officer must hold a P3K license and activity book from the local manpower agency, with requirements that include a P3K training certificate. The number follows Annex I of the regulation:
| Workplace classification | Number of workers | Number of P3K officers |
|---|---|---|
| Low hazard potential | 25 to 150 | 1 person |
| Low hazard potential | more than 150 | 1 person for every 150 persons or fewer |
| High hazard potential | 100 or fewer | 1 person |
| High hazard potential | more than 100 | 1 person for every 100 persons or fewer |
The management must also arrange for P3K officers to be available at work units 500 meters or more apart, on every floor of a multistory building, and on every shift. The duties of a P3K officer are to carry out P3K, maintain P3K facilities, record every P3K activity in the activity book, and report to the management. The names and locations of P3K officers are posted where they are easy to see.
The regulation does not state whether a hotel is low or high hazard potential, and does not set the number of officers for workplaces with fewer than 25 workers. The classification is set by the management together with the labor inspector. P3K facilities include a P3K room, first aid boxes and their contents, evacuation and transport equipment, and personal protective equipment and special equipment such as showers and eyewash stations for special hazards. A P3K room is required if the company employs 100 persons or more, or fewer than 100 persons with high hazard potential.
The regulation does not mention AEDs. An Indonesian regulation requiring AEDs in hotels cannot yet be confirmed; as of November 2026 none had been found in the JDIH databases that could be opened. For AED maintenance, the reference points are the device manufacturer’s manual and other countries’ guidance, such as the UK guidance in the following sections.
The AED: a device that guides its user
An AED (automated external defibrillator) is a device applied to the outside of the body through electrodes that gives an electric shock when a person’s heart stops pumping blood. The UK Department for Education’s AED guidance explains that once attached, the device analyzes the heart rhythm automatically, then delivers a shock if required or advises that resuscitation be continued. Modern devices give voice and visual prompts from the moment they are opened, and are designed to be used by people without specific training by following those prompts. The guidance also stresses that emergency services are called first, because help on the spot only buys time until medical staff arrive.
For a facility manager, AED readiness is therefore determined by location, reach time, and the condition of the device: enough battery, electrodes within date and connected, and no warning signs. All of that is facility work.
A self-test is an automatic check that an AED runs periodically; if it finds a problem, the device shows a warning sign or light. A self-test does not read the dates on spare packaging and does not check the cabinet, signs, or location of the device, so checks by a person are still needed.
HSE notes that where an employer in the UK provides an AED, the work equipment regulations apply, and the employer provides information and written instructions on how to use it, for example from the AED manufacturer. In a hotel, those short instructions are posted on the cabinet.
Placement, cabinets, and signs
The UK guidance asks that AEDs be located strategically so that they can be reached quickly in an emergency, in a highly visible and well signposted location. Ideally a device is no further than a two-minute brisk walk from the places where it is most likely to be needed. The location assessment considers the time needed to climb stairs, open doors, and unlock the cabinet, slip or trip hazards along the route, lighting, and the risk of tampering.
In a hotel, the locations usually considered are the security post staffed 24 hours a day, the fitness center, the swimming pool, the ballroom, and the main kitchen and laundry. The opening case shows that a busy lift can double the retrieval time, so that time is measured by actually walking it, not estimated from a floor plan.
The UK guidance gives several points for cabinets and signs:
- An indoor AED can be kept in a wall bracket or cabinet of the right size for the device, and wherever it is, it must be visible and signposted.
- An outdoor AED is kept in a thermostatically controlled cabinet in a well-lit place, because electrodes may not function as effectively if their temperature falls below about 0 °C. Such a cabinet requires a power supply and installation by a suitably qualified person.
- Cabinets and brackets are marked with the standard AED sign.
- Lockable or alarmed cabinets deter tampering, but security is weighed against the need to get the device quickly. Cabinets inside buildings generally do not require a lock.
- If an AED is temporarily moved other than in an emergency, for example for an event, a clear notice is displayed at its usual location with the phone number of the person holding it.
The minimum AED specification in the same guidance requires that the device can be stored and operated at 0 to 40 °C. That figure is a UK purchasing requirement, not a property of every device, but a metal cabinet in tropical sun can become far hotter than 40 °C. The device’s temperature range is read from its manual, and outdoor cabinets are placed in the shade.
Routine checks
The UK guidance asks for a procedure to check AEDs for warning signs regularly, no less frequently than weekly, preferably by a designated person, with a method of recording when the check took place. The condition of the cabinet, including the door closure and lock, is also checked regularly. Many AEDs require additional monthly or annual checks according to their manual, and a fault that cannot be solved with the manual during the warranty period is reported to the manufacturer immediately. AEDs do not normally require routine servicing by a technician.
The building technician’s weekly check can include:
- The device is in its cabinet, the cabinet opens, and the direction signs and AED sign are visible and lit.
- The status indicator shows ready and there is no warning light or sound.
- The adult electrodes are connected to the device, the packaging is intact, and the expiry date is recorded.
- Pediatric electrodes or a pediatric mode key, if the device requires them, are kept with the device.
- The accessories are complete: scissors, gloves, a towel or dry wipes, and a disposable razor.
- Outdoor cabinet: the temperature inside, the light, the heater or cooler, and the alarm work.
The UK guidance suggests that adult electrodes be pre-connected to the device so that it is ready faster and because many devices need this for their self-tests, while pediatric electrodes are not recommended for pre-connection because most cardiac arrests occur in adults. The technician does not open the casing and does not press buttons other than those named in the manual for checking.
Batteries, electrodes, and expiry dates
According to the UK guidance, electrodes, razors, gloves, and face shields are replaced every time the device is used, and some manufacturers also advise replacing the battery after use even if the indicator does not yet show it as low. Even when no incident has occurred, batteries and electrodes have a limited service life and are replaced after the period set by the manufacturer, usually when the expiry date on each item is reached, or for batteries when the indicator shows low, whichever comes first. Replacement items must be the correct ones for that device, because items for different AEDs are generally not compatible with each other.
The same guidance suggests keeping a minimum stock of spare electrodes and disposable accessories on site. An AED also has a service life estimated by the manufacturer, which is not necessarily the same as its warranty period, and the device is replaced when that life is reached or when its batteries and electrodes are no longer available. Software updates follow the latest resuscitation guidelines and are arranged by the manufacturer.
An expiry date is the date after which the manufacturer no longer guarantees that an item will work as intended. For first aid boxes, HSE asks that the contents of first aid containers be examined frequently, restocked soon after use, and items that reach their expiry date disposed of safely; containers of sterile water or saline whose seal has been broken are not kept for reuse, and containers are not used beyond their expiry date. In an Indonesian first aid box, items that usually carry an expiry date include sterile gauze, saline solution, povidone-iodine, alcohol, and plasters.
One list of dates for each AED and each box makes the work easier, and replacement orders are scheduled from the nearest expiry date minus the supplier’s lead time.
First aid box contents under Permenakertrans PER.15/MEN/VIII/2008
A first aid box must be made of strong, easily carried material, with a white base color and a green P3K symbol, contain the items in Annex II, and must not be filled with materials or equipment other than those needed for P3K at the workplace. Boxes are placed where they are easy to see and reach, with clear direction signs, enough light, and easy to lift, in the numbers set by Annex III; work units 500 meters or more apart and work units on different floors each provide boxes according to their number of workers.
The box contents under Annex II, written as printed in the text that could be opened:
| No | Item | Box A (25 workers or fewer) | Box B (50 workers or fewer) | Box C (100 workers or fewer) |
|---|---|---|---|---|
| 1 | Sterile wrapped gauze | 20 | 40 | 40 |
| 2 | Bandage (5 cm wide) | 2 | 4 | 6 |
| 3 | Bandage (10 cm wide) | 2 | 4 | 6 |
| 4 | Adhesive tape (1.25 cm wide) | 2 | 4 | 6 |
| 5 | Quick plasters | 10 | 15 | 20 |
| 6 | Cotton wool (25 grams) | 1 | 2 | 3 |
| 7 | Triangular bandage/sling | 2 | 4 | 6 |
| 8 | Scissors | 1 | 1 | 1 |
| 9 | Safety pins | 12 | 12 | 12 |
| 10 | Disposable gloves | 2 | 3 | 4 |
| 11 | (pairs) | 2 | 4 | 6 |
| 12 | Face mask | 1 | 1 | 1 |
| 13 | Tweezers | 1 | 1 | 1 |
| 14 | Flashlight | 1 | 1 | 1 |
| 15 | Eyewash cup | 1 | 2 | 3 |
| 16 | Clean plastic bag | 1 | 1 | 1 |
| 17 | Aquades (100 ml saline solution) | 1 | 1 | 1 |
| 18 | Povidone-iodine (60 ml) | 1 | 1 | 1 |
| 19 | 70% alcohol | 1 | 1 | 1 |
| 20 | Workplace P3K guidebook | 1 | 1 | 1 |
| 21 | Notebook | 1 | 1 | 1 |
Below row 21, “List of box contents” is also printed, without a number or quantity. Row 11 contains only the word “(pairs)” with quantities of 2, 4, and 6, which is most likely a continuation of row 10 or part of a shifted row. The arrangement of rows 10 and 11 in the text that could be opened cannot yet be confirmed; the interpretation is asked of the labor inspector or the local manpower office, and in the meantime the box is filled with the larger quantity of the two rows.
Annex III sets the number of boxes per work unit: fewer than 26 workers, 1 box A; 26 to 50 workers, 1 box B or 2 boxes A; 51 to 100 workers, and every 100 workers, 1 box C, or 2 boxes B, or 4 boxes A, or 1 box B and 2 boxes A. Its notes state that 1 box B is equivalent to 2 boxes A, and 1 box C is equivalent to 2 boxes B.
Oral medicines, ointments, and other drugs are not listed in Annex II, so they are not put in the first aid box. HSE guidance in the UK also recommends that tablets and medicines not be kept in the first aid container. For comparison, the UK has no mandatory list of contents and first aid containers are marked with a white cross on a green background; in Indonesia the contents and color of the box are set by regulation.
Record keeping and the role of trained officers
Records are the evidence that the AED and first aid boxes are ready. For each AED, the facility register holds the serial number, location, electrode and battery expiry dates, weekly checks with the checker’s name, findings and actions, and every use. For each first aid box, the register holds the location, box type, check dates, items restocked, and items discarded because they expired. Annex II of the Indonesian regulation itself includes a notebook in every box, and the P3K officer records every P3K activity in their activity book.
After an AED has been used, the UK guidance asks the facility to contact the local ambulance service to arrange for the device’s recorded data to be downloaded, to take care that the data is not erased, and then to make the device ready for use again by replacing electrodes, batteries, and disposable accessories as required and making sure no warning light or message is showing. In Indonesia, who downloads the data is arranged with the device supplier and the hotel’s referral health facility.
The division of roles:
| Party | Role |
|---|---|
| Building technician | cabinets, brackets, signs, lighting, power supply and temperature of outdoor cabinets; weekly checks and the register; ordering the correct consumables |
| Licensed P3K officer | P3K treatment, maintaining first aid box contents, the activity book, reports to management |
| Company management | appointing officers, setting the hazard classification, providing facilities, posting officers’ names and locations |
| AED supplier or manufacturer | repairs, batteries, software, recorded data, device service life |
| Emergency medical services | care of the casualty up to and after arrival at the hospital |
During an incident, the person who finds the casualty calls a trained P3K officer and emergency medical services, the person nearest the cabinet brings the AED, and whoever uses the AED follows the device’s voice prompts until medical staff arrive.
Worked example
The figures in the examples below are practice figures, unless their origin is stated.
Example 1: number of P3K officers per shift. A hotel employs 320 workers, and management has classified it as low hazard potential. For the whole workforce, Annex I gives 1 person for every 150 persons or fewer, that is, officers. Because officers must also be available on every shift, the count is made per shift: the morning shift of 160 workers needs , the afternoon shift of 110 workers needs 1, and the night shift of 50 workers needs 1, so at least licensed officers are needed, not counting cover for leave and officers per floor.
Example 2: number of first aid boxes per work unit. A housekeeping unit of 80 workers on one floor falls in the 51 to 100 range, so it needs 1 box C, or 2 boxes B, or 4 boxes A, or 1 box B and 2 boxes A. With the equivalence 1 B = 2 A, the last option is equivalent to boxes A. The main kitchen with 40 workers falls in the 26 to 50 range and needs 1 box B or 2 boxes A.
Example 3: auditing the contents of box B. Box B in the kitchen holds 25 sterile gauze pads, 6 quick plasters, 1 expired povidone-iodine, a packet of headache tablets, and a tube of ointment. The shortfalls are gauze and quick plasters . The povidone-iodine is replaced, while the headache tablets and ointment are removed because they are not listed in Annex II, and every action is recorded in the register.
Example 4: AED retrieval time. Brisk walking speed is 1.5 m/s, and opening the cabinet takes 15 seconds. From the fitness center, the AED cabinet in the lobby is 120 m away including stairs, so the one-way time is seconds and the time until the device is in hand is seconds, under two minutes, that is, 120 seconds. From the edge of the swimming pool the distance is 210 m: seconds, plus 15 makes 155 seconds, more than 120 seconds. Counted as a round trip, even from the fitness center it is seconds. This finding becomes a proposal for a second AED near the pool and fitness center.
Example 5: scheduling an electrode order. On November 15, 2026, the connected electrodes expire on March 31, 2027, and the supplier’s lead time is 6 weeks, that is, days. From November 15, 2026 to March 31, 2027 there are days. The order deadline is March 31 minus 42 days, that is, February 17, 2027, or days from the day of the check. That date goes into the register and a reminder.
Example 6: reading the check register and the cabinet temperature. Over 52 weeks, the register records 47 weekly checks, so weeks were missed and compliance is percent. Three of those were consecutive, so the longest interval between checks is weeks, long enough for electrodes to pass their date unnoticed. The outdoor cabinet near the pool recorded 46 °C at 14:00, that is, °C above the 40 °C limit in the UK minimum specification. The technician compares this with the temperature range in the device manual and reports to the supplier before moving the cabinet into the shade.
Stop conditions
- The AED status indicator shows a warning, does not show ready, or the device sounds a warning: post an “AED not ready” notice with the location of the nearest ready AED, report to the P3K officer and the duty manager, and contact the supplier; do not open the casing or try to repair it.
- The electrodes have expired or their packaging is damaged, or the battery is past its date or low: replace them only with the correct items for that device from the supplier; if there is no spare, the device is declared not ready as in item 1. Do not use electrodes or batteries from a different type of AED.
- The AED is cracked, wet, exposed to liquid, smells burnt, or its battery is swollen: do not switch on, charge, or open the device; keep it away from flammable materials and hand it to the supplier or manufacturer.
- Someone collapses or is injured: call a trained P3K officer and emergency medical services, bring the AED to the scene, and follow the device’s voice prompts. If the casualty may still be in contact with an electrical source, do not touch them until the source has been disconnected by an authorized person according to Chapter 75 · Electrical & Control; if chemicals are involved, follow Chapter 6. Other medical treatment is left to trained officers.
- After the AED has been used: do not erase or reset the device’s data; hand the device to the P3K officer and the supplier for data download and replacement of consumables, and post a notice until the device is ready again.
- There is a request to put medicines or items outside Annex II into a first aid box, or to lock the AED cabinet or first aid box so that the duty officer cannot open it: refuse, and hand the decision to the management and the P3K officer.
- The heater, cooler, light, or power supply of an outdoor AED cabinet fails: do not open live panels or cables; hand it to an authorized electrical technician, and move the AED to the nearest indoor location with a notice giving its location.
Common misinterpretations
Guess what’s wrong first, then open the explanation.
Common misconception
As long as the AED light is green, the device does not need checking.
Why is it wrong?
A self-test does not check the spare electrodes, cabinet, signs, or location of the device, and a light that nobody looks at tells nobody anything. Weekly checks and the register are still needed.
Common misconception
A complete first aid box also contains headache tablets.
Why is it wrong?
Annex II of the Indonesian regulation does not list oral medicines, and the box must not be filled with items other than those needed for P3K.
Common misconception
The AED cabinet should be locked so that it is not stolen.
Why is it wrong?
A lock adds retrieval time. Cabinets inside buildings generally do not need a lock; where security is needed, an alarm on the cabinet slows retrieval less.
Common misconception
Electrodes that were used only briefly can be used again.
Why is it wrong?
Electrodes and disposable accessories are replaced every time the device is used, and the device is declared ready only after replacement and a check of the indicator.
Common misconception
The technician who maintains the AED is thereby a P3K officer.
Why is it wrong?
A P3K officer is an appointed, licensed worker with a training certificate. Maintaining the device is a facility task, not permission to give first aid.
Links to other chapters
Chapter 75 · Electrical & Control on electrical hazards to the body and Chapter 6 on chemicals contain first aid for their respective hazards. The R23 chapter on fire evacuation drills in hotels (Chapter 14) refers to this chapter for casualties during an evacuation, and Chapter 17 · Mechanical & Mechatronics on hotel fitness equipment covers a place that often becomes an AED location.
Exercises and worked solutions
The figures in the exercises are for practice.
Try to answer first, then open the worked solution.
A low hazard potential hotel employs 140 workers on the morning shift, 60 on the afternoon shift, and 30 on the night shift. How many P3K officers must at least be available per shift?
Check my answer
See the worked solution
The morning shift needs officer because 140 is in the 25 to 150 range, the afternoon shift of 60 workers needs 1, and the night shift of 30 workers needs 1. So at least one licensed officer is present on every shift, plus officers per floor if work units are spread over several floors.
A laundry has 45 workers on the basement floor. What first aid boxes does it need?
Check my answer
See the worked solution
The 26 to 50 worker range needs 1 box B or 2 boxes A. Because the laundry is on its own floor, its boxes are provided in that unit, not borrowed from another floor.
AED electrodes expire on June 30, 2027, and the supplier’s lead time is 4 weeks. When is the order deadline?
Check my answer
See the worked solution
Four weeks is 28 days, and June 30 minus 28 days is June 2, 2027. That order deadline is recorded in the register so that new electrodes arrive before the connected ones expire.
A weekly check finds the AED indicator glowing red on the evening before a large event in the ballroom. What does the technician do?
Check my answer
See the worked solution
The device is declared not ready, a notice with the location of the nearest ready AED is posted, and the P3K officer and duty manager are informed. The technician contacts the supplier for a replacement or loan device, and does not open the casing or try to repair the device.
A department head asks for the first aid box in their office to be locked because its contents are often taken. How does the technician respond?
Check my answer
See the worked solution
Under the regulation, a first aid box must be easy to see and reach, so a locked box fails when it is needed. The problem of missing items is handled with a contents register and periodic checks by the P3K officer, and the decision is handed to the management.
Why does an AED in an outdoor cabinet in Indonesia need different attention to temperature from the example in the UK guidance?
Check my answer
See the worked solution
The UK guidance discusses heated cabinets because electrodes may be less effective below about 0 °C, while in Indonesia the problem is more often heat. A metal cabinet in the sun can exceed the 40 °C limit in the UK minimum specification, so the temperature is recorded and compared with the device manual.
References
External sources
- Peraturan Menteri Tenaga Kerja dan Transmigrasi Republik Indonesia Nomor PER.15/MEN/VIII/2008 tentang Pertolongan Pertama pada Kecelakaan di Tempat Kerjajdih.kemnaker.go.id
- Automated External Defibrillators (AEDs): Guidance for schoolsassets.publishing.service.gov.uk
- First aid at work: The Health and Safety (First-Aid) Regulations 1981. Guidance on Regulations (L74)hse.gov.uk
Revision history
| Date | Version | Change |
|---|---|---|
| November 15, 2026 | 1 | First published |
How to cite
Kurnia (2026). AEDs and first aid boxes from the facility side: placement and signs, checking batteries and electrodes, expiry dates, first aid box contents, record keeping, and the role of trained officers. ARSTEMA Academy. Revised November 15, 2026. arstema.com/en/academy/konsep/aed-dan-kotak-p3k-dari-sisi-fasilitas-penempatan-dan-tanda-pemeriksaan-baterai-dan-elektroda-masa-kedaluwarsa-isi-kotak-p3k-pencatatan-serta-peran-petugas-terlatih
Self-check · 3 questions
Check your understanding before moving on
Write your answer in your own words, then compare it with the worked solution from this chapter’s exercises. The system doesn’t judge open answers as right or wrong; you decide which parts you’ve understood.
A low hazard potential hotel employs 140 workers on the morning shift, 60 on the afternoon shift, and 30 on the night shift. How many P3K officers must at least be available per shift?
A laundry has 45 workers on the basement floor. What first aid boxes does it need?
AED electrodes expire on June 30, 2027, and the supplier’s lead time is 4 weeks. When is the order deadline?